How to Choose the Measuring Range: What 1250 / 2000 / 3000 / 6000μm / 15mm Each Map To

How to Choose the Measuring Range: What 1250 / 2000 / 3000 / 6000μm / 15mm Each Map To

The bottom line

The basis for range selection is the actual distribution of thickness, not the “largest possible value”. Selecting a range by an upper limit you will never reach costs you overall accuracy.

Scenarios mapped to each band

Range band Typical application Dongru model
0-500μm Ultra-thin paint, oxide layers on aluminium and copper DR370B
0-1250μm Thin coatings, anodising, PCB conformal coatings, automotive paint DR260A / DR280A / DR370A
0-1500μm Automotive paint, general coatings DR120 / DR130
0-2000μm General industrial coatings, galvanising, automotive body panels DR220 / DR230 / DR320 / DR330 / DR360A / DR380A
0-3000μm High-build coatings, thicker anti-corrosion layers DR3000
0-5000μm Pipeline anti-corrosion, thick coatings DR5000S
0-6000μm Heavy anti-corrosion, thick anti-corrosion layers, some fireproofing DR6000A
0-10000μm Very-wide-range fireproofing and anti-corrosion DR9000A
100μm-15mm Intumescent fireproofing and other very thick coatings DR15A

Why “running at the limit” is a bad habit

Accuracy expressions carry a proportional term, for example the 3%H in ±(1%~3%)H+1.5μm grows linearly with the reading. The closer the reading to the upper range limit, the larger the absolute tolerance. Beyond that, near the limit the probe output signal approaches saturation and non-linear error begins to appear.

How to determine the range you actually need

  1. Profile your measured thickness data over the past period (at least a few dozen samples);
  2. Take the 80th percentile as the “typical upper thickness”, not the maximum;
  3. Make that value fall in the 50%-80% band of the chosen range;
  4. Keep a plan for extreme values — measure them with another wide-range gauge rather than widening the usual band.

The relationship between units and range

Range switching is often accompanied by unit switching. US engineering practice favours mil (thousandths of an inch), 1 mil = 25.4 μm. When selecting, confirm the gauge supports μm / mil dual-unit switching, and watch the rounding direction in conversion — when stating a range outward you should round down, never overstate.

Special considerations for wide-range scenarios

  • Rougher surfaces on thick coatings: wide-range scenarios (fireproofing, heavy anti-corrosion) have high surface roughness and require base-metal zero correction;
  • Possible internal layering: multi-pass thick coatings may have inter-layer interfaces, giving larger reading fluctuation, so increase the number of measurement points;
  • Probe form: thick coatings often use a large-diameter probe for stability, but large probes adapt poorly to curved surfaces.

Common misconceptions

  • ❌ “Buy big, not small, it measures anything anyway” — covering a small-range scenario with a wide-range gauge wastes accuracy noticeably.
  • ❌ “Over-range is just slightly inaccurate” — past the upper limit the reading enters a non-linear region and the number is worthless.
  • ❌ “The lower range limit does not matter” — the lower limit is equally critical; for example a 0.55mm ultrasonic lower limit decides whether thin walls can be measured.